== P1 data collection ==
== P1 data collection / Pilatus ==


According to the classical paper ([http://dx.doi.org/10.1107/S0907444999008367 Z. Dauter (1999), Acta Cryst D55, 1703]), the required rotation range for native data in space group P1 is 180°, and for anomalous data is 180° + 2 theta_max (theta is the diffraction angle). In the case of the standard geometry (direct beam vertical to, and central upon, the detector), this leads to 2-fold redundancy. <br /> However, experience shows that collection of a few degrees more than that is a good idea, as the scaling will be more stable. So we regularly collect 200° for native data. <br /> The Pilatus 6M detector at the SLS is composed of many panels, and therefore has horizontal and vertical dead areas. This generally lowers completeness, and the effect is particularly noticeable in P1. Untested idea: it may be good to make sure (if necessary, by moving the Pilatus) that the direct beam is not at a crossing between horizontal and vertical dead areas, nor at the middle of a panel (because this avoids that equivalent reflections suffer the same fate).
According to the classical paper ([http://dx.doi.org/10.1107/S0907444999008367 Z. Dauter (1999), Acta Cryst D55, 1703]), the required rotation range for native data in space group P1 is 180°, and for anomalous data is 180° + 2 theta_max (theta is the diffraction angle). In the case of the standard geometry (direct beam vertical to, and central upon, the detector), this leads to 2-fold redundancy. <br /> However, experience shows that collection of a few degrees more than that is a good idea, as the scaling will be more stable. So we regularly collect 200° for native data. <br /> The Pilatus 6M detector at the SLS is composed of many panels, and therefore has horizontal and vertical dead areas. This generally lowers completeness, and the effect is particularly noticeable in P1. Untested idea: it may be good to make sure (if necessary, by moving the Pilatus) that the direct beam is not at a crossing between horizontal and vertical dead areas, nor at the middle of a panel (because this avoids that equivalent reflections suffer the same fate).

Revision as of 00:03, 14 December 2007

P1 data collection / Pilatus

According to the classical paper (Z. Dauter (1999), Acta Cryst D55, 1703), the required rotation range for native data in space group P1 is 180°, and for anomalous data is 180° + 2 theta_max (theta is the diffraction angle). In the case of the standard geometry (direct beam vertical to, and central upon, the detector), this leads to 2-fold redundancy.
However, experience shows that collection of a few degrees more than that is a good idea, as the scaling will be more stable. So we regularly collect 200° for native data.
The Pilatus 6M detector at the SLS is composed of many panels, and therefore has horizontal and vertical dead areas. This generally lowers completeness, and the effect is particularly noticeable in P1. Untested idea: it may be good to make sure (if necessary, by moving the Pilatus) that the direct beam is not at a crossing between horizontal and vertical dead areas, nor at the middle of a panel (because this avoids that equivalent reflections suffer the same fate).